PRIMACY 2 - Firmware update and ribbon calibration

Modified on Wed, 13 Mar 2024 at 10:19 AM

Firstly download the latest firmware from the link below:

If updating the firmware alone does not resolve the issue, then try syncing the ribbon sensors.

Navigate to the 'Support tools' section


In the 'command' box we want to type in Sa;r then hit 'Execute Command' and the Answer box should read 'OK' if it is successful.

Please note the Sa;r command is case sensitive, the S must be uppercase.

Once you have run this, do the exact same again, but this time, type Sa;c and execute that.

This will recalibrate the ribbon sensors, and in tandem with the firmware update should resolve many ribbon issues for the Primacy 2.

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select atleast one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article